
TVT2 (5/02)
Rev. 1.7 (.pdf)
21
functionalities you will be prompted to set the delay times and the set point as with Alarm
#1.
Once you complete the selections for Alarm #2 you will be prompted to set up the analog
output.
8.6 Setting the Analog Output (O/P)
The output (O/P) selection allows you to turn the 4-20 mA analog output on, or off. You
can select the desired analog output operation using the t and
u
buttons. Once you
have set the desired operation, press the
↵
button to accept it.
If you selected to turn the 4-20 mA output on, you will be prompted to set the upper
(OHV) and lower (OLV) turbidity levels corresponding to the 4 mA and 20 mA output
levels. First, you will be prompted with the turbidity level assigned to the 4 mA output
level:
Select the turbidity level you wish to assign to the OLV using the t and
u
buttons.
Once you have set the desired level, press the
↵
button to accept it.
Next, you will be prompted with the turbidity level assigned to the 20 mA output level
(OHV on the lower row of the LCD display).
Komentarze do niniejszej Instrukcji