OWNER’S MANUAL
HF scientific– Micro 1000 Laboratory
Turbidimeter 0-10,000 NTU
HF scientific
3170 Metro Parkway
Ft. Myers, FL 33916
Phone: (239) 337-2116
Fax: (239) 332-7643
Toll free: 888-203-7248
Email: HF [email protected]
Website: www.hfscientific.com
Catalog No. 22615 (1/11)
Rev 3.2
Przeglądanie stron 0
1 2 3 4 5 6 ... 32 33

Podsumowanie treści

Strona 1 - Turbidimeter 0-10,000 NTU

OWNER’S MANUAL HF scientific– Micro 1000 Laboratory Turbidimeter 0-10,000 NTU HF scientific 3170 Metro Parkway Ft. M

Strona 2

Rev. 3.2 4 The ON/OFF key ( ) key is used to turn the instrument ON or OFF. The enter key ( ) is u

Strona 3 - DECLARATION OF CONFORMITY

Rev. 3.2 5 4. Routine Operation 4.1. Measurement Principle & Methodology This instrument is d

Strona 4

Rev. 3.2 6 During normal operation, the AUTO block will be highlighted with the current scale disp

Strona 5

Rev. 3.2 7 5. Completely fill the rinsed cuvette (from step 4) with the remaining portion (approxi

Strona 6

Rev. 3.2 8 4.5. Dry Air Purge System Cold samples, or the continuous flow of cool water through t

Strona 7 - SPECIFICATIONS

Rev. 3.2 9 Under normal conditions, re-calibration is recommended at least once every three months1

Strona 8 - 2. Overview

Rev. 3.2 10 1. Slowly rotate the calibration standard one complete revolution (360°). While rotating

Strona 9

Rev. 3.2 11 3. When the instrument has completed the calibration, the “Cal” and the “Store” blocks w

Strona 10 - 3. Safety

Rev. 3.2 12 6. Insert the 100.0 NTU calibration standard into the sample well and calibrate as befor

Strona 11 - 4. Routine Operation

Rev. 3.2 13 Notes: 1. You must use the 0.02 NTU standard at the beginning and end of the calibratio

Strona 13

Rev. 3.2 14 This feature is designed to decrease the sensitivity of the instrument to dissolved

Strona 14 - 5. Calibration Procedures

Rev. 3.2 15 The upper row of the display will indicate the operational status of the security access

Strona 15

Rev. 3.2 16 This feature allows you to select the function of the serial port on the instrument. Th

Strona 16

Rev. 3.2 17 Select the desired number of days between scheduled calibrations by pressing either the

Strona 17

Rev. 3.2 18 Select the correct day by pressing either the S or T keys to change the displayed day.

Strona 18

Rev. 3.2 19 You may also reach this point by pressing either the ◄ or ► key until the word “END” app

Strona 19

Rev. 3.2 20 7. Troubleshooting 7.1. System Warning Message(s) The instrument generates warning mess

Strona 20

Rev. 3.2 21 7.1.3. Other warning messages (W-XX) The following table outlines the general warning

Strona 21

Rev. 3.2 22 WARNING ASSOCIATED MEANING TYPICAL CAUSE W-01 Lamp Failure The lamp is no longer work

Strona 22

Rev. 3.2 23 7.2. System Error Message(s) The instrument generates error messages when problems are d

Strona 23

Rev. 3.2 i DECLARAT

Strona 24

Rev. 3.2 24 8.1.Cuvette Cleaning and Care Proper measurement of the turbidity of a sample requires th

Strona 25

Rev. 3.2 25 9. Contacting the HF scientific, inc. Technical Service Department For technical assista

Strona 26 - 7. Troubleshooting

Rev. 3.2 26 11. Glossary Formazin Turbidity Units (FTU): see Nephelometric Turbidity Units Indexing a

Strona 27

Rev. 3.2 27 12. Warranty HF scientific, inc., as vendor, warrants to the original purchaser of the i

Strona 29 - 8. Routine Maintenance

Rev. 3.2 ii 1. SPECIFICATIONS...

Strona 30

Rev. 3.2 iii 10. Accessories and Replacement Parts List...

Strona 31

Rev. 3.2 1 1. SPECIFICATIONS MICRO 1000 MICRO 1000IR Instrument Conformance Conforms to specificati

Strona 32 - Glossary

Rev. 3.2 2 2. Overview The Micro 1000 process turbidimeter allows you to measure the turbidity of

Strona 33 - Warranty

Rev. 3.2 3 2.2.User Interface The user interface of the instrument is designed with a 6 key touch