MICRO 200 BW 0-10 & 0-100 NTU (05/05)
Rev. 3.4
1.3 SPECIFICATIONS FOR THE MICRO 200 BW
Specification MICRO 200 BW
Ranges: 0 - 10.0000 NTU & 0 - 100.00 NTU
Method: EPA approved* Nephelometric
Accuracy±: 10 NTU range 1% of reading or 0.005 NTU (whichever is greater)
100 NTU range 2% of reading 1 - 40 NTU, 5% of reading 40 - 100 NTU
Repeatability: 1%
Linearity
+: 1%
Resolution:
0 to 10 NTU 0.0001
1 to 100 NTU 0.01
Digital Display: LCD 6 digits
Analyzer Graphics: Built in LCD, Graphics recorder
Clock Graphics: Date and Time
Response Time: 1 to 8 seconds
Averaging Time (Electronic): 1, 16, 48 & 96 seconds, sliding average
Microprocessor: Motorola MC68HC11
Keyboard data entry system: 8 interacting membrane switches with tactile feedback
Security Code: Prevents unauthorized access to analyzer
Built in diagnostics: Yes
Analog Output, Isolated: Bracketed to .01 NTU, 4 - 20 mA & .1, 1 & 10 VDC 12 bit resolution D/A
Alarms: 1 system alarm, 2 (NTU or % T) alarms (Hi, Lo and Off)
Alarm Contact rating: Max. 250 VAC @ 5.0 A
Operating Temperature: 32° to 122°F (0°C to 50°C)
Storage Temperature: -4°F to 140°F (-20° to 60°C)
Optical Cleaning for NTU Sensor: Removable cuvette
Flow rate: 0.5 - 1.5 gpm (2000 ml/min - 6.0 L/m)
Positive System Pressure: 60 psi maximum (414 kPa or 4.22 kg/cm
3
)
Wetted surfaces: Nylon, C.A.B., PVC, Delrin-500, Borosilicate glass, Silicone
Lamp: External, prefocused module
Standard Cable Length: 6 feet (1.83 m)
Sensor to analyzer distance: 250 feet (76.25 m) max. Over 50' lengths contact factory.
Dimensions
Analyzer: 13" x 11½” x 8" (330 x 290 x 200 mm)
NTU Sensor: 8¾” x 9¾” x 8" (225 x 250 x 200 mm)
NTU Sensor Case: NEMA 4X
CE Rating: Pollution Degree 2, Overvoltage Category III
Supply Voltage: 120 VAC
+ 10% 50/60 Hz
240 VAC
+ 10% 50/60 Hz
Power Consumption: 40 VA
*EPA approved means that the instrument meets or exceeds the design and performance criteria as specified in the
United States Environmental Protection Agency Method 180.1.