Page!1DRT-100B!!(5/00)
I. SPECIFICATIONS
_________________________________________________________________________
0 - 1
Ranges NTU 0 - 10
0 - 100
0 - 1000*
_________________________________________________________________________
Linearity
+ 1% of full scale
Repeatability
+ 1% of full scale
Sensitivity .01 NTU
_________________________________________________________________________
Power Consumption 45 Watts (Nominal)
Fuse-0.5A (120V)
Fuse-0.25A (240V)
_________________________________________________________________________
Supply Voltage 120VAC±10% 50/60Hz
240VAC±10% 50/60Hz
_________________________________________________________________________
Controls - Range Switch
- Reference Adjust
- Alarm Setting
- On/Off Switch
- 120V/240V Switch
_________________________________________________________________________
Dimensions 8"W x 10 1/2"L x 8 1/2"H
_________________________________________________________________________
Shipping Weight 12 pounds
_________________________________________________________________________
Operating temperature 0° - 50°C (32° - 122°F)
_________________________________________________________________________
*NOTE: On the maximum range of 0 - 1000 NTU the Digital Meter will read as high as 1,999. Readings over 1,000
may not be within published specifications for linearity.