DRT-15CE (5/10)Page 11
9.2 LAMP REPLACEMENT
The lamp source (HF catalog #21084) used in the DRT-15CE has an
extremely long life. Before replacing the lamp, make certain that the battery
is not in need of a charge and is not defective. Prior to replacing the lamp,
remove the battery as described in battery replacement. Remove the lamp
wires at terminal block. To remove the lamp, loosen the two set screws on
the outermost barrel with a 5/64" hex wrench and pull the lamp straight
out. Replace the lamp in the reverse order. Make sure the lamp is pushed
all the way in. The set screws should be snugged up; excess pressure could
damage the lamp. Replace the battery. Replace the chassis in the case and
recalibrate as described in CALIBRATION PROCEDURES SECTION
7.0, page 5 in this manual.
10.0 PARTS & ACCESSORIES
CAT. NO. DESCRIPTION
60013 ProCal Calibration Standard 0.02 NTU (replace
annually)
21805 Instruction Manual DRT-15CE
50051 Cuvettes - Screw Top, 3/pk.
21084 Lamp Source Assembly 2/pk.
70008 Battery - 6 volt
70050 Battery Charger - 120 volt (US style)
20850 Photo Diode
19046 Battery Charger, 240 Volt (European style)
21904 LCD Display
20495 Printed Circuit Board Complete
50040 FormazinStockSolutionKit
20956 Fuse, 1A, 5 x 20 mm
39071 ProCal Calibration Standards -1 ea.10, 100, 1000
NTU, (Replace annually)
19045 Auto Adapter 12V
50091 Anti-fog kit
HFscientic
3170 Metro Parkway
Fort Myers, FL 33916-7597
Phone: (239) 337-2116
Fax: (239) 332-7643
Toll free: 888-203-7248
HFscientic
PARTS AND ACCESSORIES
FOR